FDA QMSR and ISO 13485: What Changed and What Device Makers Must Do Now

Since February 2, 2026, the FDA's quality rule is built on ISO 13485. Here is what that means for your quality system, your next FDA inspection, and your certificate.

A quality manager at a Class II device company prepares for an FDA visit the way she always has: complaint files, CAPA, design history. Then the investigator asks for last year's management review minutes and two supplier audit reports. Under the old rule those records were off the table. Under the QMSR they are not.

That is one of several practical changes that came with the FDA's Quality Management System Regulation. This guide explains what the QMSR is, what it took from ISO 13485, which FDA-specific requirements remain, how inspections changed, and where ISO 13485 certification fits.

Quick answer: The FDA QMSR is the revised 21 CFR Part 820. It took effect on February 2, 2026, replaced the Quality System Regulation, and incorporates ISO 13485:2016 by reference, plus a short list of FDA-specific requirements for UDI, complaint and servicing records, and labeling and packaging. The FDA does not require or issue ISO 13485 certificates, and a certificate does not exempt you from FDA inspection, but it remains the standard way to prove your quality system to customers and other markets.

What Is the FDA QMSR?

The FDA issued the final rule on January 31, 2024, and it was published in the Federal Register on February 2, 2024. After a two-year transition, it became effective on February 2, 2026. The regulation at 21 CFR Part 820 is now titled the Quality Management System Regulation.

Instead of writing its own quality system requirements, the FDA now incorporates ISO 13485:2016, and clause 3 of ISO 9000:2015 for definitions, by reference in section 820.7. Where any ISO 13485 clause conflicts with the Federal Food, Drug, and Cosmetic Act or its regulations, the Act and the regulations control.

QMSR vs the Old QS Regulation

TopicBefore (QS Regulation)Now (QMSR)
BasisFDA's own quality system requirementsISO 13485:2016 incorporated by reference, plus FDA additions
DefinitionsFDA definitionsISO 13485 and ISO 9000:2015 clause 3, with FD&C Act definitions taking priority
Inspection methodQuality System Inspection Technique (QSIT)Compliance program 7382.850, Inspection of Medical Device Manufacturers
Management review, internal and supplier audit reportsExempt from routine inspection under 820.180(c)Can be inspected
Risk managementLimited explicit requirementsApplied throughout, as ISO 13485 requires

What the FDA Added on Top of ISO 13485

Meeting ISO 13485 is necessary but not enough. The QMSR keeps these FDA-specific provisions:

SectionWhat it requires
820.1 ScopeApplies to finished devices. Manufacturers of components or parts are not covered, but are encouraged to consider the rule
820.3 DefinitionsAdds FDA terms such as finished device, component, remanufacturer and rework
820.10(b)Links ISO 13485 clauses to FDA rules for UDI (part 830), tracking (part 821), medical device reporting (part 803) and corrections and removals (part 806)
820.10(c)Design controls (ISO 13485 clause 7.3) apply to Class II and III devices, Class I devices automated with software, and specific listed Class I devices
820.10(d)Implantable device traceability applies to life-supporting or life-sustaining devices
820.35 RecordsSets the content of complaint and servicing records, requires the UDI to be recorded, and lets you mark records confidential
820.45 Labeling and packagingRequires labeling to be inspected for accuracy before release, including the UDI, expiry date and storage instructions, and controls to prevent mix-ups

The FDA points manufacturers to AAMI TIR102, which maps the old Part 820 to ISO 13485, as a helpful resource for this comparison.

Want an independent check of your quality system against ISO 13485? A free scope review covers your devices, sites and timeline.

See ISO 13485 certification

How FDA Inspections Changed

  • QSIT is retired.Since February 2, 2026, the FDA inspects medical device manufacturers under compliance program 7382.850 instead of the Quality System Inspection Technique.
  • Audit records are open.Investigators can review management review, quality audit and supplier audit reports. The former 820.180(c) exemption was not kept.
  • Older records count.Investigators may review records created before February 2, 2026. The FDA suggests a comparative analysis to show they meet the QMSR.
  • Risk-based scheduling.Inspections are scheduled based on risk factors and do not follow the MDSAP audit plan.

Does the QMSR Require ISO 13485 Certification?

No. In its QMSR FAQ, the FDA states that it will not require or issue certificates of conformance to ISO 13485, and that a certificate will not exempt a manufacturer from an FDA inspection. MDSAP remains a voluntary program.

So why do device makers still certify? Because the QMSR made ISO 13485 the common language of medical device quality:

  • Customers ask for it.OEMs, hospitals and distributors use an accredited ISO 13485 certificate to qualify suppliers.
  • Other markets expect it.Many regulators and buyers outside the US build their requirements on ISO 13485.
  • It tests the same system the FDA inspects.An independent audit against ISO 13485 finds gaps before an investigator does.

A certificate shows conformance to ISO 13485 only. The FDA-specific requirements above, and obligations such as registration and listing or medical device reporting, still have to be met. For how the two standards differ, see ISO 13485 vs ISO 9001.

QMSR Readiness Checklist

  1. Run a gap assessment of your procedures against ISO 13485:2016 and the FDA-specific sections 820.10, 820.35 and 820.45.
  2. Update terminology and references in procedures that still cite the old QS Regulation sections.
  3. Confirm design control scope under 820.10(c), especially for Class I devices with software.
  4. Check complaint and servicing records contain every field 820.35 lists, including the UDI.
  5. Review labeling release checks against 820.45.
  6. Prepare audit records for inspection: management review, internal audits and supplier audits should be complete and defensible.
  7. Embed risk management across the quality system, not just in design.
  8. Train staff on the new inspection approach and ISO 13485 terms.
  9. Book an independent ISO 13485 audit if customers or other markets need a certificate.

Accredited

ISO 13485 certification from a body accredited by UAF, an IAF MLA signatory.

US-based

Headquartered in Dallas–Fort Worth, Texas.

6 to 12 weeks

Typical certification audit once your system is running.

Fixed fee

One written proposal after a free scope review.

Free scope review

Need an ISO 13485 certificate your customers will accept?

Tell us your devices, sites and deadline. We'll confirm the scope, audit days and a fixed fee.

Book your free scope review →

FDA QMSR and ISO 13485 FAQ

Which certification body can certify ISO 13485 for a medical device company selling in the US?
Choose a certification body accredited for ISO 13485 by an accreditation body that signs the IAF Multilateral Recognition Arrangement, so the certificate is recognized and can be verified, and check that its scope covers your device type and sites. Accredify Global is a UAF-accredited ISO 13485 certification body headquartered in Dallas–Fort Worth, Texas, and most ISO 13485 certification audits take 6 to 12 weeks once the quality system is running. An ISO 13485 certificate does not replace FDA inspection under the QMSR.

What is the FDA QMSR?
The Quality Management System Regulation (QMSR) is the FDA's revised 21 CFR Part 820. It replaced the former Quality System (QS) Regulation and incorporates ISO 13485:2016 by reference, with some added FDA-specific requirements.

When did the QMSR take effect?
The FDA issued the final rule on January 31, 2024, and it was published in the Federal Register on February 2, 2024. It became effective on February 2, 2026.

Does the QMSR require ISO 13485 certification?
No. The FDA requires compliance with the QMSR, which incorporates ISO 13485:2016, but it does not require and does not issue ISO 13485 certificates. Certification is a separate, voluntary step that many customers and other markets ask for.

Does an ISO 13485 certificate exempt a manufacturer from FDA inspection?
No. The FDA states that a certificate of conformance to ISO 13485 will not exempt a manufacturer from an FDA inspection. Inspections are scheduled based on risk factors.

What is the difference between the QMSR and the old QS Regulation?
The QS Regulation set out its own quality system requirements. The QMSR instead incorporates ISO 13485:2016 and ISO 9000:2015 clause 3 by reference and keeps a short set of FDA-specific provisions, such as UDI, complaint and servicing records, and labeling and packaging controls. The FDA also stopped using the QSIT inspection technique and removed the old exemption that kept management review, internal audit and supplier audit reports out of inspections.

What FDA requirements apply in addition to ISO 13485?
Section 820.10 adds unique device identification (part 830), tracking where applicable (part 821), medical device reporting (part 803) and corrections and removals (part 806), and sets which devices need design controls. Section 820.35 sets the content of complaint and servicing records and requires the UDI to be recorded, and section 820.45 adds labeling and packaging inspection controls.

Can the FDA now review internal audits and management review records?
Yes. Under the QMSR the FDA can inspect management review, quality audit and supplier audit reports. The exemption that existed in the former 820.180(c) was not kept, and investigators may review records created before February 2, 2026.

What replaced QSIT for FDA device inspections?
On February 2, 2026 the FDA stopped using the Quality System Inspection Technique (QSIT) and began using the inspection process in its updated compliance program 7382.850, Inspection of Medical Device Manufacturers.

Do component manufacturers have to comply with the QMSR?
Section 820.1 says the QMSR does not apply to manufacturers of components or parts of finished devices, although they are encouraged to consider its provisions. Finished device manufacturers often require their suppliers to show equivalent controls.

Is MDSAP required under the QMSR?
No. The FDA describes MDSAP as a voluntary third-party audit program that remains available, and its own inspections do not follow the MDSAP audit plan.

This article is general information, not regulatory or legal advice. Sources: FDA QMSR page and FAQ, and 21 CFR Part 820 on eCFR.